British journal of anaesthesia
-
Review Meta Analysis
Accuracy and precision of non-invasive cardiac output monitoring devices in perioperative medicine: a systematic review and meta-analysis†.
Cardiac output (CO) measurement is crucial for the guidance of therapeutic decisions in critically ill and high-risk surgical patients. Newly developed completely non-invasive CO technologies are commercially available; however, their accuracy and precision have not recently been evaluated in a meta-analysis. We conducted a systematic search using PubMed, Cochrane Library of Clinical Trials, Scopus, and Web of Science to review published data comparing CO measured by bolus thermodilution with commercially available non-invasive technologies including pulse wave transit time, non-invasive pulse contour analysis, thoracic electrical bioimpedance/bioreactance, and CO2 rebreathing. ⋯ Inter-study sensitivity heterogeneity was high (I2=83%, P<0.001). With a wide percentage error, completely non-invasive CO devices are not interchangeable with bolus thermodilution. Additional studies are warranted to demonstrate their role in improving the quality of care.