Thromb Haemostasis
-
Multicenter Study
A multicenter evaluation of a new quantitative highly sensitive D-dimer assay for exclusion of venous thromboembolism.
D-dimer testing is widely applied for exclusion of deep-vein thrombosis (DVT) and pulmonary embolism (PE). We report on a multicenter performance evaluation of a new particle-enhanced immunoassay, Innovance D-Dimer. Innovance D-Dimer assay was performed in 1,543 frozen samples from outpatients suspected of DVT and/or PE enrolled in three management studies as well as in a routine clinical practice. ⋯ Two samples from patients with distal DVT tested negative with all assay systems. One patient with high pre-test clinical probability and proximal DVT tested negative with Vidas D-Dimer Exclusion. Our data indicate that the performances of Innovance D-Dimer, regardless of the analyzer, are similar to the reference methods, and that this assay can be used for the exclusion of venous thromboembolic disease.