Singap Med J
-
Randomized Controlled Trial
Does pulse oximetry accurately monitor a patient's ventilation during sedated endoscopy under oxygen supplementation?
Pulse oximetry (SpO2) measures oxygen saturation but not alveolar ventilation. Its failure to detect alveolar hypoventilation during sedated endoscopy under oxygen supplementation has been reported. The aim of this study was to measure the masking effect of oxygen supplementation in SpO2 when alveolar hypoventilation develops during sedated endoscopy. ⋯ SpO2 alone is not adequate for monitoring alveolar ventilation during sedated endoscopy under oxygen supplementation due to possible delays in detecting alveolar hypoventilation in patients. Even if SpO2 decreases by only 1% during the procedure and its level remains near 100%, physicians should consider the onset of severe alveolar hypoventilation, which requires immediate intervention.