Respiratory care
-
Multicenter Study
DLCO Biologic Quality-Control Findings From a Multi-Center Global Study.
The 2017 American Thoracic Society/European Respiratory Society (ATS/ERS) diffusing capacity of the lung for carbon monoxide (DLCO) standards specify a control rule for assessing biologic quality control (BioQC) but have limited guidance on how to establish expected values for control rule variables. This study aimed to determine expected values for DLCO BioQC using coefficient of variation (CV) and compare that the mean ± 2 SD control rule yields the same precision as mean ± 12% of the mean. ⋯ A DLCO BioQC CV ≤ 6% is achievable across multiple sites, technologists, and brands of equipment. This CV value assures that measurements for control rule variables emerge from an expected range. A control rule of mean ± 2 SD appeared to yield similar results as the mean ± 12% of the mean rule reported in the 2017 ATS/ERS DLCO standards.