Physics in medicine and biology
-
Comparative Study
Ultrasound elastic tensor imaging: comparison with MR diffusion tensor imaging in the myocardium.
We have previously proven the feasibility of ultrasound-based shear wave imaging (SWI) to non-invasively characterize myocardial fiber orientation in both in vitro porcine and in vivo ovine hearts. The SWI-estimated results were in good correlation with histology. In this study, we proposed a new and robust fiber angle estimation method through a tensor-based approach for SWI, coined together as elastic tensor imaging (ETI), and compared it with magnetic resonance diffusion tensor imaging (DTI), a current gold standard and extensively reported non-invasive imaging technique for mapping fiber architecture. ⋯ The average ETI-estimated fractional anisotropy (FA) values decreased from subendocardium to subepicardium (p < 0.05, unpaired, one-tailed t-test, N = 10) by 33%, whereas the corresponding DTI-estimated FA values presented a change of -10% (p > 0.05, unpaired, one-tailed t-test, N = 10). In conclusion, we have demonstrated that the fiber orientation estimated by ETI, which assesses the shear wave speed (and thus the stiffness), was comparable to that measured by DTI, which evaluates the preferred direction of water diffusion, and have validated this concept within the myocardium. Moreover, ETI was shown capable of mapping the transmural fiber angles with as few as seven shear wave propagation directions.