Physics in medicine and biology
-
The aim of this paper is to describe a new automatic method for compensation of metal-implant-induced segmentation errors in MR-based attenuation maps (MRMaps) and to evaluate the quantitative influence of those artifacts on the reconstructed PET activity concentration. The developed method uses a PET-based delineation of the patient contour to compensate metal-implant-caused signal voids in the MR scan that is segmented for PET attenuation correction. PET emission data of 13 patients with metal implants examined in a Philips Ingenuity PET/MR were reconstructed with the vendor-provided method for attenuation correction (MRMap(orig), PET(orig)) and additionally with a method for attenuation correction (MRMap(cor), PET(cor)) developed by our group. ⋯ MRMap(cor) only showed small residual segmentation errors in eight patients. ε(rel)(av) (mean ± standard deviation) were: (-56 ± 3)% for B1, (-43 ± 4)% for B2, (21 ± 18)% for L1, (120 ± 47)% for L2 regions. ε(rel) (mean ± standard deviation) of hot focal structures were: (-52 ± 12)% in B1, (-45 ± 13)% in B2, (19 ± 19)% in L1, (51 ± 31)% in L2 regions. Consequently, metal-implant-induced artifacts severely disturb MR-based attenuation correction and SUV quantification in PET/MR. The developed algorithm is able to compensate for these artifacts and improves SUV quantification accuracy distinctly.