Biomedical instrumentation & technology / Association for the Advancement of Medical Instrumentation
-
Biomed Instrum Technol · Nov 1998
Treatment of out-of-hospital cardiac arrest with a low-energy impedance-compensating biphasic waveform automatic external defibrillator. The LIFE Investigators.
Few victims of sudden cardiac arrest survive. A new generation of automatic external defibrillators (AEDs), smaller, lighter, easier to use, and less costly, makes the goal of widespread AED deployment and early defibrillation feasible. A low-energy impedance-compensating biphasic waveform allows AED device characteristics more suitable to the goal of early defibrillation than high-energy waveforms. ⋯ At the time of patient transfer, an organized rhythm was present in 65% of the VF patients; asystole was the result in 25%, and VF was in progress in 10%. It is concluded that low-energy impedance-compensating biphasic waveforms terminate long-duration VF at high rates in out-of-hospital cardiac arrest and provide defibrillation rates exceeding those previously achieved with high-energy shocks. Use of this waveform allows AED device characteristics consistent with widespread AED deployment and early defibrillation.