Mannequin-assessed dry-heat exchanges in the incubator-nursed newborn.
For manufacturers of warming devices for newborn infants, knowledge of the partition of the various channels of heat transfer between the neonate and the environment is necessary to regulate them adequately. The goal of this study was to determine the contributions of the different components of dry heat exchange using a thermal mannequin that replicated a full-term neonate. Each mannequin segment could be controlled separately at a selected surface temperature. ⋯ Whatever the incubator temperature, a fivefold increase in convective heat exchange was observed when air velocity increased, whereas radiative heat exchange was unchanged. At an incubator temperature of 34 degrees C and for air velocities between 0.1 and 0.4 m/sec radiative heat exchange is the dominant mode of heat loss. The results suggest that this thermal mannequin is a good tool for assessing dry heat exchange in incubators.