Seminars in thrombosis and hemostasis
-
Semin. Thromb. Hemost. · Apr 2015
Comparative StudyInternational normalized ratio monitoring of vitamin K antagonist therapy: comparative performance of point-of-care and laboratory-derived testing.
The monitoring of warfarin therapy using the international normalized ratio (INR) has now moved outside the laboratory's control by use of point-of-care (POC) devices. Although this provides patients with the convenience of immediate results and clinical assessment, POC-INRs are often performed by nonlaboratory staff with little experience in quality control. The Royal College of Pathologists of Australasia Quality Assurance Program (RCPAQAP) Haematology has devised a POC-INR external quality assessment (EQA) program that is suitable for both laboratory and nonlaboratory operators (e.g., nurses) to perform INR testing with good accuracy and precision. ⋯ A comparative assessment with other EQA providers showed agreement and disparity with our data in terms of comparative CVs obtained between the laboratory and POC-INRs. The growth of the RCPAQAP POC-INR program from 29 to 360 in the past 12 years highlights the importance of providing suitable EQA for POC-INR staff who are unfamiliar with laboratory practice. This helps maintaining consistent results, which have important implications for the therapeutic management of patients on vitamin K antagonist therapy.