Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
-
Multicenter Study Clinical Trial
Safety, efficacy, and performance of new discrimination algorithms to reduce inappropriate and unnecessary shocks: the PainFree SST clinical study design.
Implantable cardioverter defibrillator (ICD) shock therapy improves survival of patients at risk for sudden cardiac death. The high sensitivity of ICDs to detect tachycardia events is accompanied by reduced specificity resulting in inappropriate and unnecessary shocks. Up to 30% of ICD patients may experience inappropriate shocks, which are most commonly caused by lead noise, oversensing of T-waves, and supraventricular tachycardias. The new Protecta ICD and cardiac resynchronization therapy devices have been designed to minimize inappropriate and unnecessary shocks through novel SmartShock(TM) technology algorithms targeting these causes. ⋯ Inappropriate and unnecessary shock delivery remains a significant clinical issue for patients receiving device therapies, which has considerable consequences for patients and the healthcare system. The PainFree SST study will investigate the ability of new algorithms to reduce inappropriate shocks. Results from this study are expected in mid-2013.