Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
-
Multicenter Study Clinical Trial
First prospective, multi-centre clinical experience with a novel left ventricular quadripolar lead.
Cardiac resynchronization therapy (CRT) is sometimes complicated by elevated pacing thresholds and phrenic nerve stimulation (PNS), both of which may require that the coronary sinus lead be repositioned. The purpose of this study was to evaluate the performance of a novel quadripolar electrode lead and cardiac resynchronization therapy-defibrillator (CRT-D) device that enables electrical repositioning, potentially obviating a lead reposition procedure. ⋯ The new quadripolar electrode LV lead provides more programming options to address common problems faced when managing CRT patients. Electrical measurements from new vectors are comparable with conventional configurations. Furthermore, 11% of patients in the study suffered PNS on all conventional bipolar vectors.