Physiological reports
-
Physiological reports · Jun 2015
Response characteristics of esophageal balloon catheters handmade using latex and nonlatex materials.
The measurement of esophageal pressure allows for the calculation of several important and clinically useful parameters of respiratory mechanics. Esophageal pressure is often measured with balloon-tipped catheters. These catheters may be handmade from natural latex condoms and polyethylene tubing. ⋯ Balloon catheters handmade from polyurethane condoms displayed the smallest working volume range and lowest static balloon compliance. Despite this lower compliance, fA 5% and fφ5% were remarkably similar between all balloon materials. Our findings suggest that polyisoprene condoms are an ideal nonlatex construction material to use when fabricating esophageal catheter balloons by hand.