Radiation medicine
-
The purpose of this study was to evaluate the accuracy of T2 calculation from single shot imaging sequences such as echo-planar imaging (EPI) and half-Fourier single shot turbo spin-echo (HASTE) imaging. ⋯ With single shot sequences, it is possible to obtain the T2 values that show excellent correlation with the CPMG sequence. Although both HASTE and EPI are useful to calculate T2 values, EPI appears to be more accurate in characterizing focal liver lesions.
-
The present prospective survey was performed to obtain information on delayed adverse reactions (DARs) to five types of low-osmolar iodinated contrast media, including their frequency, common manifestations, and the patient's history of allergy. ⋯ Four risk factors for DARs were identified in the present investigation.
-
A real-time interactive MR imaging system (real-time MRI) is an MR scanner which has a fast image updating cycle and the ability to freely change slice orientation, just like an ultrasound imaging system. Recently, such a system has been developed and installed on a clinical 1.5-Tesla system. The purpose of this study was to optimize the pulse sequences for clinical use and to evaluate the clinical usefulness and basic functionality of real-time MRI. ⋯ Maximum C/N between cardiac cavity and wall was obtained at a flip angle of 60 degrees on true FISP. Localization of the right and left coronary arteries could be performed within 30 seconds in three volunteers. Although the present real-time MRI system has drawbacks such as low spatial resolution and relatively low contrast resolution, we expect real-time MRI to be one of the most important tools for future clinical MRI.