J. Cardiothorac. Vasc. Anesth. · Feb 2021
Observational StudyComparing the Mutual Interchangeability of ECOM, FloTrac/Vigileo, 3D-TEE, and ITD-PAC Cardiac Output Measuring Systems in Coronary Artery Bypass Grafting.
- Adam J Milam, Farhad Ghoddoussi, Jon Lucaj, Spurthy Narreddy, Nakul Kumar, Vennela Reddy, Joffer Hakim, and Sandeep H Krishnan.
- Department of Anesthesiology, Cedars-Sinai Medical Center, Los Angeles, CA.
- J. Cardiothorac. Vasc. Anesth. 2021 Feb 1; 35 (2): 514-529.
ObjectiveThe aim of this study was to compare the mutual interchangeability of 4 cardiac output measuring devices by comparing their accuracy, precision, and trending ability.DesignA single-center prospective observational study.DesignNonuniversity teaching hospital, single center.ParticipantsForty-four consecutive patients scheduled for elective, nonemergent coronary artery bypass grafting (CABG).InterventionsThe cardiac output was measured for each participant using 4 methods: intermittent thermodilution via pulmonary artery catheter (ITD-PAC), Endotracheal Cardiac Output Monitor (ECOM), FloTrac/Vigileo System (FLOTRAC), and 3-dimensional transesophageal echocardiography (3D-TEE).Measurements And Main ResultsMeasurements were performed simultaneously at 5 time points: presternotomy, poststernotomy, before cardiopulmonary bypass, after cardiopulmonary bypass, and after sternal closure. A series of statistical and comparison analyses including ANOVA, Pearson correlation, Bland-Altman plots, quadrant plots, and polar plots were performed, and inherent precision for each method and percent errors for mutual interchangeability were calculated. For the 6 two-by-two comparisons of the methods, the Pearson correlation coefficients (r), the percentage errors (% error), and concordance ratios (CR) were as follows: ECOM_versus_ITD-PAC (r = 0.611, % error = 53%, CR = 75%); FLOTRAC_versus_ITD-PAC (r = 0.676, % error = 49%, CR = 77%); 3D-TEE versus ITD-PAC (r = 0.538, % error = 64%, CR = 67%); FLOTRAC_versus_ECOM (r = 0.627, % error = 51%, CR = 75%); 3D-TEE_versus ECOM (r = 0.423, % error = 70%, CR = 60%), and 3D-TEE_versus_FLOTRAC (r = 0.602, % error = 59%, CR = 61%).ConclusionsBased on the recommended statistical measures of interchangeability, ECOM, FLOTRAC, and 3D-TEE are not interchangeable with each other or to the reference standard invasive ITD-PAC method in patients undergoing nonemergent cardiac bypass surgery. Despite the negative result in this study and the majority of previous studies, these less-invasive methods of CO have continued to be used in the hemodynamic management of patients. Each device has its own distinct technical features and inherent limitations; it is clear that no single device can be used universally for all patients. Therefore, different methods or devices should be chosen based on individual patient conditions, including the degree of invasiveness, measurement performance, and the ability to provide real-time, continuous CO readings.Copyright © 2020 Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.