• Spine · Dec 2020

    Chromium Metal Ion Release During Instrumented Spinal Surgery In Children; The Effects Of Electrosurgery.

    • Peter J Cundy, Georgia Antoniou, Annika Mascarhenas, Freeman Brian J C BJC Department of Orthopaedic Surgery, Women's and Children's Hospital, Adelaide, South Australia, Australia. , and William J Cundy.
    • Department of Orthopaedic Surgery, Women's and Children's Hospital, Adelaide, South Australia, Australia.
    • Spine. 2020 Dec 1; 45 (23): 1619-1624.

    Study DesignProspective 2-year study with operative sampling and in-vitro analysis of chromium ions following spinal surgery in children.ObjectivesTo measure metal ion levels at preoperative, intraoperative, and postoperative times to determine patterns of metal ion release during instrumented spinal surgery.Summary Of Background DataRaised serum metal ion levels are reported following instrumented spinal fusion in adolescent idiopathic scoliosis. The authors noted raised chromium levels in patients receiving implants that did not contain chromium. This prompted further work to establish the source. The electrosurgery tips used were discovered to contain 17% chromium.MethodsChromium, cobalt, and titanium levels were measured before, during, and after surgery in serum and local intraoperative fluid samples in 11 children undergoing posterior instrumentation for scoliosis. Administered drugs, cell-saver fluids, and intraoperative fluids, both local and intravenous, were investigated to exclude these as sources of chromium ions. An in-vitro study was also performed to elucidate sources of intraoperative chromium ions.ResultsHigh chromium levels were detected in all samples from the wound irrigation fluid prior to insertion of metal implants. Immediate postoperative chromium serum ion levels were also elevated and returned to baseline by day 30. In-vitro sampling of fluids from test models using electrosurgery revealed high levels of chromium ions CONCLUSION.: This finding of high chromium metal ion concentrations in intraoperative and early postoperative samples provides evidence of chromium release during the dissection phase of spinal surgery. This challenges existing beliefs that metal ion release occurs solely due to implants and now implicates the electrosurgery electrode tips as a source of raised chromium ion levels. Thorough irrigation of the operative site after the dissection phase of surgery to both dilute and reduce the intraoperative chromium ion load is suggested. Alternative electrosurgery electrode tips or other methods to coagulate during surgery could be considered.Level Of Evidence2.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.