• Arch Orthop Trauma Surg · Aug 2015

    Is there a rationale to use a dual mobility poly insert for failed Birmingham metal-on-metal hip replacements? A retrieval analysis.

    • Lisa Renner, Martin Faschingbauer, and Friedrich Boettner.
    • Hospital for Special Surgery, 535 East 70th Street, New York, NY, 10021, USA, rennerl@hss.edu.
    • Arch Orthop Trauma Surg. 2015 Aug 1;135(8):1177-81.

    IntroductionPrevious studies showed poor outcomes for patients undergoing revision of failed metal-on-metal total hip arthroplasty (MoM-THA) and resurfacing (RS) with an increased risk of dislocation. Dual mobility inserts are an option to retain the acetabular component and change the metal-on-metal bearing to plastic-on-metal. The current study analyzes the rationale for the off-label use of a dual mobility poly insert (MDM X3, Stryker, Mahwah, NJ) in a Birmingham metal shell (Smith & Nephew, Memphis, TN).Materials And MethodsBased on retrievals from the implant database the study compared the clearance between 20 BHR shells, 31 MDM poly inserts and 24 ADM acetabular components of different sizes. The radial clearance was calculated for each possible combination of implants [n = 81 (MDM/BHR) and n = 119 (MDM/ADM)].ResultsAn MDM mobile bearing poly insert in an ADM shell has an average clearance of 0.314 mm (SD 0.031) compared to 0.234 mm (SD 0.030) in a BHR shell (p < 0.01). The minimal clearance is 0.246 and 0.163 mm, respectively. 30.9 % of the MDM/BHR clearances were within the range of the MDM/ADM bearing and 88.9 % had a clearance of more than 0.2 mm.ConclusionClearances of the MDM poly insert in a BHR shell are reduced, and although the majority of combinations appear safe, the indication needs to be made on an individual base carefully considering alternative treatment options.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.