• Spine · Apr 2024

    Are all cages created equal? Analysis of Cervical Cage Malfunctions Using FDA MAUDE Database.

    • Victor Ungurean, Amit S Piple, Oluwatodimu Richard Raji, Andrea Rowland, Adam Schlauch, Dimitriy G Kondrashov, Ken Y Hsu, and James F Zucherman.
    • The Taylor Collaboration, San Francisco, CA.
    • Spine. 2024 Apr 15; 49 (8): 553560553-560.

    Study DesignRetrospective case series.ObjectiveTo characterize failure rates of cervical cages based on manufacturer and design characteristics using the nationwide database of reported malfunctions.BackgroundThe Food and Drug Administration (FDA) aims to ensure the safety and efficacy of cervical interbody implants postimplantation; however, intraoperative malfunctions may be overlooked.Materials And MethodsThe FDA's Manufacturer and User Facility Device Experience database was queried for reports of cervical cage device malfunctions from 2012 to 2021. Each report was categorized based on the failure type, implant design, and manufacturer. Two market analyses were performed. First, "failure-to-market share indices" were generated by dividing the number of failures per year for each implant material by its yearly US market share in cervical spine fusion. Second, "failure-to-revenue indices" were calculated by dividing the total number of failures per year for each manufacturer by their approximate yearly revenue from spinal implants in the US. Outlier analysis was performed to generate a threshold value above which failure rates were defined as greater than the normal index.ResultsIn total, 1336 entries were identified, and 1225 met the inclusion criteria. Of these, 354 (28.9%) were cage breakages, 54 (4.4%) were cage migrations, 321 (26.2%) were instrumentation-related failures, 301 (24.6%) were assembly failures, and 195 (15.9%) were screw failures. Poly-ether-ether-ketone implants had higher failure by market share indices for both migration and breakage compared with titanium. Upon manufacturer market analysis, Seaspine, Zimmer-Biomet, K2M, and LDR exceeded the failure threshold.ConclusionThe most common cause of implant malfunction was breakage. Poly-ether-ether-ketone cages were more likely to break and migrate compared with titanium ones. Many of these implant failures occurred intraoperatively during instrumentation, which underscores the need for FDA evaluation of these implants and their accompanying instrumentation under the appropriate loading conditions before commercial approval.Copyright © 2023 Wolters Kluwer Health, Inc. All rights reserved.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…