    In vitro model of prepacked carbon dioxide absorber use: development and testing.

    • Jennifer Jouwena, Delphine Verbeke, Andre M De Wolf, Arne Neyrinck, and HendrickxJan F AJFA0000-0002-1360-9662Department of Basic and Applied Medical Sciences, Ghent University, Ghent, Belgium; Department of Anesthesiology, UZLeuven, Leuven, Belgium; Department of Cardiovascular Sciences, KULeuven, Leuven, Belgium; Department.
    • Department of Anesthesiology, Onze Lieve Vrouw Hospital, Aalst, Belgium; Department of Anesthesiology, UZLeuven, Leuven, Belgium; Department of Cardiovascular Sciences, KULeuven, Leuven, Belgium.
    • Anesthesiology. 2024 Mar 1; 140 (3): 450462450-462.

    BackgroundCarbon dioxide absorbers allow the use of fresh gas flow below minute ventilation (V˙E). Models are developed and tested in vitro to quantify their performance with variable carbon dioxide load (V˙CO2), fresh gas flow, V˙E, end-tidal carbon dioxide (ETco2) fraction, and the type of workstation used.MethodsFirst principles are used to derive a linear relationship between fresh gas flow and fractional canister usage or FCU0.5 (the reciprocal of the time for the inspiratory carbon dioxide fraction to reach 0.5%). This forms the basis for two basic models in which V˙E was measured by spirometry or calculated. These models were extended by multiplying V˙E with an empirical workstation factor. To validate the four models, two hypotheses were tested. To test whether the FCU0.5 intercept varied proportionally with V˙CO2 and was independent of V˙E, FCU was measured for 10 canisters tested with a fixed 0.3 l/min fresh gas flow and a range of V˙CO2 while V˙E was either constant or adjusted to maintain ETco2 fraction. A t test was used to compare the two groups. To confirm whether a change in V˙CO2 accompanied by a change in V˙E to maintain ETco2 fraction would shift the linear fresh gas flow-FCU0.5 relationship in a parallel manner, 19 canisters were tested with different combinations of V˙CO2 and fresh gas flow. These measured FCU values were compared to those predicted by the four models using Varvel's performance criteria.ResultsWith 0.3 l/min fresh gas flow, FCU0.5 was proportional with V˙CO2 and independent of whether V˙E was adjusted to maintain ETco2 fraction or not (P = 0.962). The hypothesized parallel shift of the fresh gas flow-FCU0.5 relationship was confirmed. Both extended models are good candidate models.ConclusionsThe models predict prepacked canister performance in vitro over the range of V˙E, fresh gas flow, and V˙CO2 likely to be encountered in routine clinical practice. In vivo validation is still needed.Copyright © 2023 American Society of Anesthesiologists. All Rights Reserved.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.