• Anesthesiology · Oct 2007

    Comparative Study

    Accuracy of cerebral monitoring in detecting cerebral ischemia during carotid endarterectomy: a comparison of transcranial Doppler sonography, near-infrared spectroscopy, stump pressure, and somatosensory evoked potentials.

    • Stefan Moritz, Piotr Kasprzak, Matthias Arlt, Kai Taeger, and Christoph Metz.
    • Department of Anesthesiology, University of Regensburg, Regensburg, Germany. stefan.moritz@klinik.uni-regensburg.de
    • Anesthesiology. 2007 Oct 1;107(4):563-9.

    BackgroundThis study compares the accuracy of cerebral monitoring systems in detecting cerebral ischemia during carotid endarterectomy.MethodsThe authors compared transcranial Doppler sonography (TCD), near-infrared spectroscopy (NIRS), stump pressure (SP) measurement, and somatosensory evoked potentials (SEP) in 48 patients undergoing carotid surgery during regional anesthesia. Cerebral ischemia was assumed when neurologic deterioration occurred. During clamping, the minimum mean middle cerebral artery velocity (TCD(min)), its percentage change (TCD%), the minimum regional saturation of oxygen (NIRS(min)), its percentage change (NIRS%), the mean SP, and the changes of SEP amplitude were recorded. To analyze the corresponding sensitivity and specificity of each parameter, the authors performed receiver operating characteristic analysis.ResultsNeurologic deterioration occurred in 12 patients. SP and NIRS were successfully performed in all patients. TCD monitoring was not possible in 10 (21%); SEP was not possible in 2 patients (4%). All parameters provided the ability to distinguish between ischemic and nonischemic patients. TCD% and NIRS% showed significantly better discrimination than TCD(min) and NIRS(min) (P < 0.05). The highest area under the curve (AUC) was found for TCD% (AUC = 0.973), but there was no significant difference compared with NIRS% (AUC = 0.905) and SP (AUC = 0.925). The lowest AUC was found for SEP (AUC = 0.749), which was significantly lower than that for TCD%, NIRS%, and SP.ConclusionsTCD%, NIRS%, and SP measurement provide similar accuracy for the detection of cerebral ischemia during carotid surgery. Lower accuracy was found for SEP monitoring. Because of the high rate of technical difficulties (21%), TCD monitoring was the least practical of the investigated monitoring devices.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…