• Br J Anaesth · Mar 2013

    Comparative Study Clinical Trial

    Stroke volume optimization in elective bowel surgery: a comparison between pulse power wave analysis (LiDCOrapid) and oesophageal Doppler (CardioQ).

    • J Nordström, C Hällsjö-Sander, R Shore, and H Björne.
    • Department of Anaesthesiology, Surgical Services and Intensive Care Medicine, Karolinska University Hospital, 171 76 Stockholm, Sweden.
    • Br J Anaesth. 2013 Mar 1;110(3):374-80.

    BackgroundGoal-directed fluid therapy improves outcome in major surgery. We evaluated a new device (LiDCOrapid) against our standard oesophageal Doppler method (ODM) for stroke volume (SV) optimization during colorectal surgery.MethodsThis was an observational study in 20 patients undergoing major colorectal surgery within a fast-track protocol. We compared SV values measured simultaneously by LiDCOrapid and ODM before and after 86 fluid challenges. We also evaluated the LiDCOrapid dynamic indices SV variation (SVV) and pulse pressure variation (PPV) as predictors for volume responsiveness, defined as an increase in SV ≥ 10% after 200 ml of colloid.ResultsSV increased ≥ 10% after 27 out of 86 fluid challenges. For 172 paired SV values, the overall correlation was r=0.39, and bias (limits of agreement) -28 (-91-35) ml, percentage error 70%. The ability of LiDCOrapid to track changes in SV was weak with a concordance rate of 80%, and a sensitivity and specificity of 48% and 81%, respectively, to detect a positive fluid challenge. The area under the curve values (with 95% confidence intervals) for SVV and PPV were 0.72 (0.60-0.83) and 0.66 (0.52-0.79), respectively, indicating low predictive capacity in these setting.ConclusionsLiDCOrapid and ODM devices are not interchangeable. We cannot recommend that the LiDCOrapid replace the standard Doppler method until further device-specific outcome studies on volume optimization are available. The dynamic indices SVV and PPV add little value to a fluid optimization protocol, and should not replace SV measurements with a validated technique.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…