• Spine · Jan 1997

    A biomechanical study of a cervical spine stabilization device: Roy-Camille plates.

    • M E Smith, M Cibischino, N A Langrana, C K Lee, and J R Parsons.
    • Department of Mechanical and Aerospace Engineering, Rutgers-The State University of New Jersey, Piscataway, USA.
    • Spine. 1997 Jan 1;22(1):38-43.

    Study DesignThree-hole Roy-Camille posterior plates (Howmedica, Inc., Rutherford, NJ) were used to fix severely destabilized fresh cadaveric cervical spines. Fixed spine constructs were tested mechanically in flexion-extension and torsion, and the results were compared with the same characteristics in the intact spine before destabilization. Stainless steel and titanium plates and screws were evaluated.ObjectivesTo determine if the application of Roy-Camille posterior plates provided suitably strong and rigid fixation of a severe, surgically created three-column instability.Summary Of Background DataControversy still remains regarding the exclusive use of posterior cervical plating in the face of three-column instability. Posterior plating has been evaluated biomechanically in severely destabilized calf spines; however, posterior plating of similarly destabilized human cadaveric cervical spines using the Roy-Camille system has not been examined.MethodsThe authors chose to test the main motions of the neck (flexion, extension, and torsion) in the intact and the plated state using a servohydraulic materials testing system. Testing the surgically altered spine before fixation proved to be futile because of drastic instability, which is characteristic of the chosen defect. Once fixed, the spines were tested, and the rigidity of the constructs were compared with that of the intact state. Strength and failure mechanisms were evaluated.ResultsThe rigidity of the plated spine constructs surpassed that of the intact spines; the stainless steel and titanium systems were mechanically equivalent. Thus, application of the Roy-Camille plates of either type dramatically reduced the motion of the unstable spine. Strength of the fixed spine constructs was limited by screw pull-out at theoretically predictable levels of force.ConclusionsPosterior application of Roy-Camille plates can fix cervical spines with severe destabilizing defects rigidly. Screw pull-out of the most proximal or distal screw was always the mechanism of failure.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.