• Anesthesia and analgesia · Nov 2011

    The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery.

    • Ba-Vinh Nguyen, Jean-Louis Vincent, Emmanuel Nowak, Michelle Coat, Nicolas Paleiron, Pierre Gouny, Mehdi Ould-Ahmed, Maité Guillouet, Charles Christian Arvieux, and Gildas Gueret.
    • Department of Anesthesiology and Critical Care Medicine, University Hospital, CHU Brest, and Université de Brest, Faculté de Médecine et des Sciences de la Santé, Laboratoire de Physiologie, Brest, France.
    • Anesth. Analg.. 2011 Nov 1;113(5):1052-7.

    BackgroundIn March 2008, a new multiwavelength pulse oximeter, the Radical 7 (Rad7; Masimo Corp., Irvine, CA), was developed that offers noninvasive measurement of hemoglobin concentration. Accuracy has been established in healthy adults and some surgical patients, but not in cardiac surgery intensive care patients, a group at high risk of postoperative bleeding events and anemia in whom early diagnosis could improve management.MethodsIn this prospective, observational study conducted in a cardiovascular intensive care unit, we compared hemoglobin concentrations shown by the Rad7 with arterial hemoglobin concentrations determined by an automated hematology analyzer, XE-2100 (Roche, Neuilly sur Seine, France). Two software versions of Rad7 (V 7.3.0.1 [42 points of comparison in 14 patients] and the updated V 7.3.1.1 [61 points of comparison in 27 patients]) were studied during two 1-week periods. Bias, defined as the difference between the 2 methods (Masimo SpHb-XE-2100 laboratory hemoglobin), was calculated. A negative bias indicated that the Masimo underestimated hemoglobin compared with the laboratory analyzer. Correlation between the perfusion index given by Rad7 and the hemoglobin bias was also studied.ResultsCorrelations between Rad7 and XE-2100 were weak for both software versions (R2=0.11 for V 7.3.0.1 and R2=0.27 for V 7.3.1.1). Mean bias was -1.3 g/dL for V 7.3.0.1 and -1.7 g/dL for V 7.3.1.1, with wide 95% prediction intervals for the bias (respectively, -4.6 to 2.1 g/dL and -5.7 to 2.3 g/dL). The absolute hemoglobin bias tended to increase when the perfusion index decreased. For the V 7.3.0.1 software, the average absolute bias was 1.9 g/dL for perfusion index<2 and 0.8 g/dL for perfusion index>2 (P=0.03). For V 7.3.1.1, the mean absolute bias was 2.1 g/dL when the perfusion index was <2, and 1.6 g/dL when the perfusion index was >2 (P=0.26).ConclusionsOur study demonstrates poor correlation between hemoglobin measured noninvasively by multiwavelength pulse oximetry and a laboratory hematology analyzer. The difference was greater when the pulse oximetry perfusion index was low, as may occur in shock, hypothermia, or vasoconstriction patients. The multiwavelength pulse oximetry is not sufficiently accurate for clinical use in a cardiovascular intensive care unit.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…