• Resp Care · Jan 2001

    An evaluation of Automode, a computer-controlled ventilator mode, with the Siemens Servo 300A ventilator, using a porcine model.

    • S J Holt, R C Sanders, T L Thurman, and M J Heulitt.
    • Critical Care Medicine, Department of Pediatrics, University of Arkansas for Medical Sciences, 800 Marshall Street, Little Rock AR 72202, USA.
    • Resp Care. 2001 Jan 1;46(1):26-36.

    BackgroundWeaning of mechanical ventilation in patients optimally includes meeting their needs by making frequent ventilator adjustments. The Siemens Servo 300A mechanical ventilator is designed to allow the ventilator to be interactive with the patient's needs by making breath-by-breath adjustments in both control and support modes. We undertook the following experiment to validate that the Automode algorithm responded appropriately using a pediatric animal model when apnea occurred and if there was any impact on work of breathing.MethodsWe ventilated 6 sedated spontaneously-breathing piglets using Automode in pressure-regulated volume control/volume support (PRVC/VS) mode, pressure control/pressure support (PC/PS) mode, and volume control/volume support (VC/VS) mode. Data were collected using both a computerized respiratory monitor and data acquisition system that recorded and analyzed individual animal breaths for response time, effort of triggering, and work of breathing. Data collection began with the animals breathing spontaneously in each support mode, followed by the administration of a short-acting neuromuscular blocker (succinylcholine) to induce apnea, thus allowing the ventilator to switch between modes automatically. Data collection was continued before, during, and after apnea to observe the duration of inspiratory effort, trigger response time, and any significant pressure or flow variances of the Automode feature. In addition, patient work of breathing (WOB(P)) and ventilator work of breathing (WOB(V)) were measured before and after each phase.ResultsWe found no instances of failure of Automode to follow the predetermined algorithms. There was a difference in both the amount of change in pressure and most negative deflection of pressure by each animal during triggering in the post-paralysis phase (p < 0.05). Response time for individual breaths was shorter from initiation of breath to most negative deflection of pressure during the post-paralysis phase (p < 0.05). Maximum flow reached was lower in the post-paralysis phase for VC/VS and PC/PS (p < 0.05). We also found WOB(P) decreased and WOB(V) increased in the post-paralysis phase for all modes tested.ConclusionsThe Automode algorithm performed as expected in this animal experiment. We conclude that differences in response time and negative deflection of pressure, as an indication of animal effort, and maximum flow reached were due to continued weakness from the neuro-muscular blocker. However, the ventilator continued to trigger despite decreased effort by the animal.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…