    In vitro performance of a new non-invasive extracorporeal saturation monitor.

    • D H Bailey, E J da Silva, and T H Clutton-Brock.
    • Consultant Anaesthetist, Alexandra Hospital, Redditch, UK.
    • Anaesthesia. 2011 Oct 1;66(10):889-94.

    AbstractA simulated cardiopulmonary bypass circuit was constructed in the laboratory to compare the accuracy and precision of a new non-invasive extracorporeal oxygen saturation monitor-- the M3 monitor (Spectrum Medical LLP, Gloucester, UK) against the Siemens RAPIDLab(®) blood gas analyser (Siemens AG, Munich, Germany). Comparisons were made across a range of oxygen saturations and at different temperatures and different haemoglobin concentrations. Results showed that under all conditions, when recording oxygen saturation, the M3 monitor achieved a mean (SD) bias of -0.9 (1.7)% with limits of clinical agreement of -4.2 to 2.5. However, the difference between the two monitors was larger at lower saturation levels (p < 0.001), lower haemoglobin levels (p = .002) and lower temperatures (p = 0.013). The mean (SD) haemoglobin concentration was 15.4 (0.6) when recorded by the M3 monitor, compared to 15.4 (0.5) g.dl(-1) measured by the blood gas analyser The M3 monitor compares favourably with the gold standard of the blood gas analyser and has the advantage of giving a continuous reading. You can respond to this article at http://www.anaesthesiacorrespondence.com.© 2011 The Authors Anaesthesia © 2011 The Association of Anaesthetists of Great Britain and Ireland.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.