- B A McGrath, S Ruane, J McKenna, and S Thomas.
- University Hospital South Manchester, Manchester, UK.
- Anaesthesia. 2017 Jan 1; 72 (1): 36-41.
AbstractDisposable bronchoscopes such as the Ambu aScopeTM 3 are marketed as 'single use' The risks of contamination from prolonged device storage before possible re-use are unknown. Following clinical bronchoscopy in patients whose lungs were mechanically ventilated, 20 aScopeTM 3's bronchoscopes received a standard 'social clean' and were then stored. Subsequent paired saline flush and swab samples were taken at time zero, and at 24 h and 48 h. Positive microbiological cultures were obtained from at least one time point from 16 of the 20 bronchoscopes. Pathogens considered at high risk of causing pneumonia were isolated from seven bronchoscopes, with significant quantities from six of them. Our study demonstrates that aScopeTM 3's should not be re-used on the same patient, as clinically significant growth of micro-organisms occurs frequently, despite adequate social cleaning. Culture of bronchoscopes themselves may be a potentially useful diagnostic tool in the context of pulmonary infection. Our data make it clear that these devices are single use and not single patient use.© 2016 The Association of Anaesthetists of Great Britain and Ireland.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.