Acta Anaesthesiol Scand · Mar 2013
Randomized Controlled TrialCan electromyographic arousal be detected visually on the Datex-Ohmeda S/5™ anesthesia monitor?
- A J Aho, A Yli-Hankala, L-P Lyytikäinen, K Kamata, and V Jäntti.
- Coxa Ltd, Hospital for Joint Replacement, Tampere, Finland. antti.j.aho@uta.fi
- Acta Anaesthesiol Scand. 2013 Mar 1;57(3):364-72.
BackgroundElectroencephalogram (EEG)-based depth of anaesthesia monitoring is susceptible to contaminating electromyographic (EMG) activity. Many authorities have suggested that anaesthesiologists using these monitors should interpret the raw EEG waveform seen on the anaesthesia monitor.MethodsIn 34 patients anaesthetized with propofol using two doses of rocuronium (0.6 and 1.2 mg/kg), we studied whether the EMG arousal can be detected visually on the anaesthesia monitor. The Bispectral Index (BIS) and Entropy biosignals on the monitor were recorded with a video camera, and the one-channel EEG recorded by the Entropy strip was collected on a laptop computer. The recordings and the one-channel EEG were analyzed offline by two experts (anaesthesiologist and neurophysiologist), both with a long experience on anaesthesia-related EEG.ResultsEMG arousal existed in 14/34 and 13/33 patients in the BIS and Entropy biosignals, respectively. The anaesthesiologist detected EMG on the monitor in 7/14 patients with BIS (sensitivity 50%) and in 4/13 patients with Entropy (31%). The clinical neurophysiologist detected EMG in 6/14 (43%) patients with BIS and in 5/13 (38%) with Entropy. The specificity of the EMG analyses was 55 and 65% with BIS, and 85 and 90% with Entropy. EMG arousal was detected in BIS biosignal in 10/17 and 4/17 patients with 0.6 and 1.2 mg/kg doses of rocuronium (P = 0.04).ConclusionsIn contrast to many EEG phenomena, EMG activity cannot be accurately detected visually from the raw EEG on the anaesthesia monitor. Further development in the quality of the anaesthesia monitors is warranted.© 2012 The Acta Anaesthesiologica Scandinavica Foundation.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.