• Anesthesia and analgesia · Oct 2018

    Clinical Trial

    Design and Evaluation of a Closed-Loop Anesthesia System With Robust Control and Safety System.

    • Nicholas West, Klaske van Heusden, Matthias Görges, Sonia Brodie, Aryannah Rollinson, Christian L Petersen, Guy A Dumont, J Mark Ansermino, and Richard N Merchant.
    • From the Departments of Anesthesiology, Pharmacology, and Therapeutics.
    • Anesth. Analg. 2018 Oct 1; 127 (4): 883-894.

    BackgroundClosed-loop control of anesthesia involves continual adjustment of drug infusion rates according to measured clinical effect. The NeuroSENSE monitor provides an electroencephalographic measure of depth of hypnosis (wavelet-based anesthetic value for central nervous system monitoring [WAVCNS]). It has previously been used as feedback for closed-loop control of propofol, in a system designed using robust control engineering principles, which implements features specifically designed to ensure patient safety. Closed-loop control of a second drug, remifentanil, may be added to improve WAVCNS stability in the presence of variable surgical stimulation. The objective of this study was to design and evaluate the feasibility of a closed-loop system for robust control of propofol and remifentanil infusions using WAVCNS feedback, with an infusion safety system based on the known pharmacological characteristics of these 2 drugs.MethodsWith Health Canada authorization, research ethics board approval, and informed consent, American Society of Anesthesiologists I-III adults, requiring general anesthesia for elective surgery, were enrolled in a 2-phase study. In both phases, infusion of propofol was controlled in closed loop during induction and maintenance of anesthesia, using WAVCNS feedback, but bounded by upper- and lower-estimated effect-site concentration limits. In phase I, remifentanil was administered using an adjustable target-controlled infusion and a controller was designed based on the collected data. In phase II, remifentanil was automatically titrated to counteract rapid increases in WAVCNS.ResultsData were analyzed for 127 patients, of median (range) age 64 (22-86) years, undergoing surgical procedures lasting 105 (9-348) minutes, with 52 participating in phase I and 75 in phase II. The overall control performance indicator, global score, was a median (interquartile range) 18.3 (14.2-27.7) in phase I and 14.6 (11.6-20.7) in phase II (median difference, -3.25; 95% confidence interval, -6.35 to -0.52). The WAVCNS was within ±10 of the setpoint for 84.3% (76.6-90.6) of the maintenance of anesthesia in phase I and 88.2% (83.1-93.4) in phase II (median difference, 3.7; 95% confidence interval, 0.1-6.9). The lower propofol safety bound was activated during 30 of 52 (58%) cases in phase I and 51 of 75 (68%) cases in phase II.ConclusionsAdding closed-loop control of remifentanil improved overall controller performance. This controller design offers a robust method to optimize the control of 2 drugs using a single sensor. The infusion safety system is an important component of a robust automated anesthesia system, but further research is required to determine the optimal constraints for these safe conditions.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…