Anesthesiology
-
The Bispectral Index (BIS) reportedly reflects anesthetic depth. It is recommended that anesthetic agents should be titrated to maintain the BIS between 40 and 60 arbitrary BIS units during anesthesia. For anesthesia providers to follow this recommendation, the monitor should be predictably affected by different anesthetic agents and have good interpatient and intrapatient reproducibility. The authors hypothesized that when two BISxp devices (Aspect Medical Systems, Newton, MA) are placed concurrently on the same patient, their readings are concordant throughout the anesthetic period. ⋯ A conflicting anesthetic management was suggested by the simultaneous BIS readings 10.7% of the time. These results suggest that BISxp does not always provide a reproducible single number. Anesthesia providers should not rely exclusively on the BIS reading when assessing depth of anesthesia.
-
Ventilator-associated pneumonia is a clear risk factor for acute lung injury which has been poorly described in trauma patients. This prospective study was undertaken to estimate the incidence of such ventilator-associated pneumonia leading to acute lung injury, the risk factors, and the associated morbidity and mortality in a group of multiple trauma patients. ⋯ Acute lung injury complicated the course of 15% of ventilator-associated pneumonia in trauma patients. H. influenzae seemed to be one of the most frequent bacteria involved and the sole risk factor identified. Occurrence of ventilator-associated pneumonia leading to acute lung injury modified the prognosis of trauma patients.