Anesthesiology
-
The accuracy and precision of currently available, widely used acceleromyograph and electromyograph neuromuscular blockade monitors have not been well studied. In addition, the normalization of the train-of-four ratio from acceleromyography (train-of-four ratio [T4/T1] divided by the baseline train-of-four ratio) has not been validated in comparison to mechanomyography. ⋯ Three currently available acceleromyograph monitors produced overshoot and substantial variability that could be clinically significant. Normalization corrected the overshoot in the average results but did not reduce the wide variability. Three electromyograph monitors measured the train-of-four ratio with minimal overshoot and variability, similar to a mechanomyograph.