Multicenter StudyClinical performance of the new cardiac markers troponin T and CK-MB on the Elecsys 2010. A multicentre evaluation.
Elecsys assays for the cardiac markers Troponin T (cTnT) and CK-MB have been evaluated in an international multicenter study on the random access analyzer Elecsys 2010 to characterize their clinical performance and their comparability with respective established routine methods. In method comparison studies of Elecsys Troponin T (TnT) with Enzymun-Test TnT, good correlations (r > or = 0.95) and a high degree of correspondence (slopes in 4 laboratories between 0.95 and 1.05) were found. The method comparison studies of Elecsys CK-MB with various CK-MB routine methods lead to good correlations but some systematic deviation in the slopes due to varying standardization. ⋯ The slightly improved sensitivity of Elecsys TnT in the lower range even allows the recognition of pathological increase at cTnT concentrations below 0.1 milligrams/l in special situations with sufficient additional clinical information. Summarizing, provide the two cardiac markers on the Elecsys 2010 at least equivalent or even superior diagnostic information in various clinical situations of cardiac disease compared with routine methods. The short turn-around time and reliable performance qualify the Elecsys assays as new methods of choice for routine and emergency use.