Review Meta Analysis Comparative StudyComparison of invasive and noninvasive blood hemoglobin measurement in the operating room: a systematic review and meta-analysis.
Noninvasive hemoglobin (Hb)-monitoring devices are new inventions in pulse oximeter systems that show hemoglobin levels continuously. The aim of this systematic review and meta-analysis was to evaluate the accuracy and precision of noninvasive versus standard central laboratory Hb measurements in the operating room. We systematically searched multiple databases. ⋯ A total of 28 studies on 2000 participants were included in the meta-analysis. Meta-analysis results of mean differences between noninvasive and the central laboratory Hb measurements in overall pooled random effects were - 0.27 (95% LoA (0.44, - 0.10); P value < 0.05). According to this meta-analysis, noninvasive hemoglobin measurement has acceptable accuracy in comparison with the standard invasive method.