Anaesthesia
-
It is important to detect and treat hypovolaemia; however, detection is particularly challenging in the conscious, spontaneously breathing patient. Eight healthy male volunteers were monitored using four minimally invasive monitors: Vigileo FloTrac(™) ; LiDCOrapid(™) ; USCOM 1A; and CardioQ(™) oesophageal Doppler. ⋯ Receiver operator characteristic curves for predicting > 10% blood loss had areas under the curve of 0.68-0.82. The minimally invasive cardiac output devices tested can detect blood loss by a reduction in stroke volume in awake volunteers, and may have a role in guiding fluid replacement in conscious patients with suspected hypovolaemia.
-
Measurement of left ventricular stroke volume and cardiac output is very important for managing haemodynamically unstable or critically ill patients. The aims of this study were to compare stroke volume measured by three-dimensional transoesophageal echocardiography with stroke volume measured using a pulmonary artery catheter, and to examine the ability of three-dimensional transoesophageal echocardiography to track stroke volume changes induced by haemodynamic interventions. This study included 40 cardiac surgery patients. ⋯ The concordance rate in the 4-quadrant analysis after phenylephrine and ephedrine administration was 75% and 84%, respectively. In the polar-plot analysis, the angular concordance rate was 66% and 73% after phenylephrine and ephedrine administration, respectively. Three-dimensional transoesophageal echocardiography was clinically acceptable for measuring stroke volume; however, it was not sufficiently reliable for tracking stroke volume changes after haemodynamic interventions.