• Spine · Mar 2016

    Fatigue Performance of Cortical Bone Trajectory Screw Compared to Standard Trajectory Pedicle Screw.

    • Yusuf Tahiri Akpolat, Serkan İnceoğlu, Nolan Kinne, Devon Hunt, and Wayne K Cheng.
    • *Orthopedic Biomechanics Laboratory, Department of Orthopedics, Loma Linda University, Loma Linda, CA†School of Medicine, Loma Linda University, Loma Linda, CA‡Spine Section, Department of Orthopedic Surgery, Loma Linda University, Loma Linda, CA.
    • Spine. 2016 Mar 1; 41 (6): E335-41.

    Study DesignCadaveric biomechanical study.ObjectiveTo determine fatigue behavior of cortical bone trajectory (CBT) pedicle screws.Summary Of Background DataCortical bone trajectory screws have been becoming popular in spine surgery; however, the long-term fatigue behavior of the new CBT screws remains understudied and limitations not well defined.MethodsTwelve vertebrae from six cadaveric lumbar spines were obtained. After bone mineral density (BMD) measurements, each vertebral body was instrumented with screws from each group, that is, CBT (4.5 × 25 mm) or standard pedicle screw (6.5 × 55 mm). A load (± 4 Nm sagittal bending) was applied under displacement control at 1 Hz. Each construct was loaded for 100 cycles or until 6° of loosening was observed. After fatigue testing, the screws were pulled out axially at 5 mm/min.ResultsThe standard pedicle screw showed better resistance against 100 cycle loading compared with the CBT screws (P < 0.001, 6.9° ± 4.8° vs. 15.2° ± 5.5°, respectively). The standard pedicle screw testing usually required more than 100 cycles of loading to achieve the critical loosening (3592 ± 4564 cycles), whereas the CBT screw never exceeded 100 cycles (84 ± 24 cycles) (P = 0.002). Increased BMD was significantly associated with a higher number of cycles and less loosening. The standard pedicle screw group had a higher postfatigue pullout load than the CBT screw group (P = 0.001, 776 ± 370 N and 302 ± 232 N, respectively).ConclusionThe standard pedicle screw had a better fatigue performance compared with the CBT screw in vertebra with compromised bone quality. The proper insertion of the CBT screw might be prevented by the laminar anatomy depending on the screw head design. The CBT screw damaged the bone along its shaft by rotating around a fulcrum, located at either the pars, pedicle isthmus, or the junction of the pedicle and superior endplate, contingent upon the strength of the bone.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.