• Anesthesiology · Nov 2018

    Comparative Study Observational Study

    Comparison of the TOFscan and the TOF-Watch SX during Recovery of Neuromuscular Function.

    • Glenn S Murphy, Joseph W Szokol, Michael J Avram, Steven B Greenberg, Torin D Shear, Mark Deshur, Jessica Benson, Rebecca L Newmark, and Colleen E Maher.
    • From the Department of Anesthesiology, NorthShore University HealthSystem, University of Chicago Pritzker School of Medicine, Evanston, Illinois (G.S.M., J.W.S., S.B.G., T.D.S., M.D., J.B., R.L.N., C.E.M.) the Department of Anesthesiology, Northwestern University Feinberg School of Medicine, Chicago, Illinois (M.J.A.).
    • Anesthesiology. 2018 Nov 1; 129 (5): 880-888.

    What We Already Know About This TopicWHAT THIS ARTICLE TELLS US THAT IS NEW: BACKGROUND:: Quantitative neuromuscular monitoring is required to ensure neuromuscular function has recovered completely at the time of tracheal extubation. The TOFscan (Drager Technologies, Canada) is a new three-dimensional acceleromyography device that measures movement of the thumb in multiple planes. The aim of this observational investigation was to assess the agreement between nonnormalized and normalized train-of-four values obtained with the TOF-Watch SX (Organon, Ireland) and those obtained with the TOFscan during recovery from neuromuscular blockade.MethodsTwenty-five patients were administered rocuronium, and spontaneous recovery of neuromuscular blockade was allowed to occur. The TOFscan and TOF-Watch SX devices were applied to opposite arms. A preload was applied to the TOF-Watch SX, and calibration was performed before rocuronium administration. Both devices were activated, and train-of-four values were obtained every 15 s. Modified Bland-Altman analyses were conducted to compare train-of-four ratios measured with the TOFscan to those measured with the TOF-Watch SX (when train-of-four thresholds of 0.2 to 1.0 were achieved).ResultsBias and 95% limits of agreement between the TOF-Watch SX and the TOFscan at nonnormalized train-of-four ratios between 0.2 and 1.0 were 0.021 and -0.100 to 0.141, respectively. When train-of-four measures with the TOF-Watch SX were normalized, bias and 95% limits of agreement between the TOF-Watch SX and the TOFscan at ratios between 0.2 and 1.0 were 0.015 and -0.097 to 0.126, respectively.ConclusionsGood agreement between the TOF-Watch SX with calibration and preload application and the uncalibrated TOFscan was observed throughout all stages of neuromuscular recovery.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.